350
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_pnoffset.fm
CALL[:CELL]:PNOFfset
CALL[:CELL]:PNOFfset
Diagram Conventions on page 248
CALL[:CELL]:PNOFfest
Function Sets/queries the pseudonoise (PN) offset. The PN offset is the timing of the cells short codes
relative to system time.
This command may take up to 2 seconds to complete. The PN offset cannot be changed while a
call is connected.
Setting Range: 0 to 511
Query Range: 0 to 511
*RST Setting 12
Programming Example
OUTPUT 714;"CALL:CELL:PNOFFSET 333" sets the PN offset to 333.
:PNOFfsetCALL
[:CELL]
<sp><num value>