S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_sync.fm
CALL:SYNC
CALL:SYNC
Diagram Conventions on page 248
CALL:SYNC[:SLEVel][:SELected]
Function The setting form of this command performs two functions: (1) sets the SYNC (synchronization
channel) level, and (2) turns the SYNC state to ON.
The query form of this command returns the SYNC level for the current system type.
The optional [:SELected] keyword in this command specifies that the SYNC level and state being
set/queried applies to the current system type. See CALL[:CELL]:SYSTem[:TYPE] to query or
change the current system type.
Setting Range: 0 to -20 dB
Resolution: 0.01 dB
Query Range: 0 to -20 dB
Resolution: 0.01 dB
*RST Setting -16 dB
Programming Example
OUTPUT 714;"CALL:SYNC:SLEVel -16"
!Sets the SYNC state to ON and sets the level to -16 dB.
<sp><num value>[DB]
? (returns num value)
:SYNC
<sp><num value>[DB]
[:SLEVel]
?(returns num value)
<sp>1|ON|0|OFF
:STATe
? (returns 1|0)
Complex Command
[:SELected]
:DIGital2000
:LEVel
CALL
[:SELected]
:DIGital2000
[:SELected]
:DIGital2000
:DIGital95
:DIGital95
:DIGital95

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals