S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_system_communicate.fm
SYSTem:COMMunicate
SYSTem:COMMunicate
Diagram Conventions on page 248
SYSTem:COMMunicate:GPIB:[:SELF]:ADDRess
Function Sets/queries the test sets GPIB address.
Setting Range: 0 to 30
Resolution: 1
Query Range: 0 to 30
Resolution: 1
Factory
setting
14 (this parameter is not affected by any reset operation and can only be changed by direct user
access)
Related Topics Configuring the Test Sets GPIB on page 774
Programming Example
OUTPUT 714;"SYSTEM:COMMUNICATE:GPIB:SELF:ADDRESS 14" !Sets the GPIB address to 14.
SYSTem :COMMunicate
:DEBug
<sp>1|ON|0|OFF
? (returns 1|0)
<sp><num value>
? (returns num value)
[:SELF]
:ADDRess
[:STATe]
:LAN
<sp><string>
? (returns string)
[:SELF]
:ADDRess
:GPIB
<sp><string>:DGATeway
<sp><string>:SMASk
? (returns string)
? (returns string)