ABORt Subsystem
ABORt Subsystem
Description
The ABORt command causes a measurement cycle in progress to stop. If the measurement is not being
continuously armed (trigger arm set to single) , the measurement will remain in the idle state after this event.
If the measurement is being continuously armed (trigger arm set to continuous), a new measurement cycle
will begin after ABORt. If an ABORt command is issued from any measurement state other than measuring,
the command is ignored.
Other Commands that Execute an ABORt Action
INITiate:<meas> will execute an ABORt:<meas> as part of the INITiate:<meas> command.
READ:<meas>? will execute an ABORt:<meas> action that aborts just one trigger sequence and then
combines the INITiate and FETCh? commands.
Syntax Diagram and Command Descriptions
ABORt

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals