S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_clpcontrol.fm
CALL[:CELL]:CLPControl
CALL[:CELL]:CLPControl
Diagram Conventions on page 248
CALL[:CELL]:CLPControl:REVerse:MODE
Function Sets/queries the reverse CDMA channel closed loop power control mode.
When ACTive is selected, the test set will send power control bits that raise or lower the power
level of the mobile station depending on the received signal strength of the reverse channel.
When UP is selected, the test set sends all 0 (up) power control bits.
When DOWN is selected, the test set sends all 1 (down) power control bits.
When ALTernate is selected, the test set sends alternating up/down power control bits.
Setting Range: ACTive|UP|DOWN|ALTernate
Query Range: ACT|UP|DOWN|ALT
*RST Setting ACTive
Programming Example
OUTPUT 714;"CALL:CELL:CLPCONTROL:REVERSE:MODE ALTERNATING"
!Sets the closed loop power control mode to alternating up, down bits.
:CLPControl
CALL
<sp>ACTive|UP|DOWN|
ALTernating
? (returns ACT|UP|DOWN|ALT
:REVerse
:MODE
[:CELL]
<sp>DB1|DBHalf|DBQuarter
? (returns DB1|DBH|DBQ)
:STEP

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals