Programming a Frequency Stability Measurement
Programming a Frequency Stability Measurement
This section provides an example of how to make analog frequency stability measurements via GPIB.
The following program example assumes that a signal within the measurements range is connected to the RF
IN/OUT connector (see Input Signal Requirements on page 57).
The basic procedure to test Frequency Stability is as follows:
1. Set the operating mode to CW (see CALL[:CELL]:OPERating ).
2. Enter the expected power (see RFANalyzer:CW:EXPected:POWer ).
3. Enter the expected frequency (see RFANalyzer:MANual:MEASurement[:MFRequency] ).
4. Define the Frequency Stability Setup parameters as desired for measurement timeout and
multi-measurements (see SETup:FSTability on page 546).
5. Use the INITiate command to start the measurement.
6. Use the FETCh:FSTability command to obtain the measurement results.
Programming Example
10 !Set the expected frequency and power.
20 OUTPUT 714;SETup:FSTability:CONTinuous OFF !Set the frequency stability
30 !measurement to single trigger mode.
40 OUTPUT 714;SETup:FSTability:TIMeout 1 S !Set a timeout value of 1 second
50 !in case the measurement cannot be made.
60 OUTPUT 714;SETup:FSTability:COUNt 5 !Set multi-measurements to 5.
70 OUTPUT 714;INITiate:FSTability !Start the measurement.
80 OUTPUT 714;FETCh:FSTability? !Query the returned integrity indicator and the
90 !Worst Case Frequency Error and average Frequency measurements.
100 ENTER 714;Integrity,Freq_error,Frequency !Enter the returned values into variables.
110 IF Integrity=0 THEN !See if the integrity value = 0 (valid measurement).
120 ELSE
130 PRINT Measurement problem; check for signal level over/under range condition.
140 END IF
150 END
Returned Values
The measurements returned by this program are:
Integrity returns the measurement integrity indicator; 0 means a successful measurement with no
errors.
Freq_error returns the worst case frequency error in parts per million (PPM). Frequency returns the
average frequency.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals