225
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_prog_gen_integrity.fm
Integrity Indicator
Example Program
10 OUTPUT 714;”INITiate:DAPower” !Start Digital Average measurement
20 OUTPUT 714;”FETCH:DAPower?” !Request measurement results.
30 ENTER 714;Integrity,Tx_power !Read measurement results.
40 IF Integrity = 0 THEN !Permits measurement to be printed if integrity indicator
50 !indicates a successful measurement
60 PRINT “Analog TX Power =”;Tx_power!if 0 then measurement was successful
70 ELSE
80 PRINT “Measurement integrity questionable, integrity value = “;Integrity !If
90 !integrity
100 !not zero
110 !then print
120 !integrity
130 !value.
140 END IF
150 END
Related Topics
“Reading Error Messages”