EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #163 background imageLoading...
Page #163 background image
163
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_prog_dap.fm
Programming a Digital Average Power Measurement
850 !***Display measurements if the mobile station did not drop the call***
860 OUTPUT A;"CALL:STATUS?"
870 ENTER A;Call_status$
880 IF Call_status$="CONN" THEN
890 PRINT "Integrity indicator = ";Integrity
900 PRINT "Maximum power = ";Avg_power_dbm;"dBm";", ";Avg_power_dbw;"dBW"
910 ELSE
920 PRINT "Mobile station dropped call, invalid results"
930 END IF
940 !
950 !***Return to default call processing states***
960 OUTPUT A;"CALL:CLPCONTROL:REVERSE:MODE ACTIVE" ! Returns to active power control bits
970 OUTPUT A;"CALL:POW:DIG2000 -50" ! Returns cell power to -50 dBm/1.23 MHz
980 !
990 DISP "Test complete"
1000 OUTPUT A;"CALL:END"
1010 END
1020 Timeout: SUB Timeout
1030 PRINT "Program timed out"
1040 CLEAR 7
1050 CLEAR 714
1060 SUBEND
Returned Values
This program returns an integrity indicator followed by a digital average power measurement result (if results are
available).
Related Topics
Digital Average Power Measurement Description on page 51
Digital Average Power Troubleshooting on page 708

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals