S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_description.fm
FETCh? Subsystem
FETCh? Subsystem
Description
The FETCh? query is a function that allows users to query results from a measurement that was previously
INITiated or READ. It does NOT begin a measurement. If no measurement is in progress it will return the
integrity and measurement values from the last measurement made, or return an integrity of No Result
Available and results of NAN. If a measurement is in process, the query will hang until the results are
available, or the measurement fails or times out. The exact results returned with a FETCh? will depend on the
specific measurement. A measurement may have a number of different results or combination of results for a
FETCh?. The FETCh? queries are intended to be used to provide overlapped operation access to measurement
results from the test set. When used along with SETup and INITiate commands , FETCh? is the primary way
for the user to retrieve measurement results. In order to use the test sets concurrent test capabilities the
overlapped commands of INITiate and FETCh? must be used. Overlapped commands allow the user to send
commands and not wait for completion.
Syntax Diagrams and Command Descriptions
FETCh:AFANalyzer on page 415 (Audio Frequency Analyzer measurements)
FETCh:ATXPower on page 424 (Analog Transmitter Power measurements)
FETCh:CAPPower on page 427 (Access Probe Power measurements)
FETCh:CCTPhase on page 428 (Code Channel Time/Phase Error measurements)
FETCh:CFERror on page 432 (Frame Error Rate test results)
FETCh:CPOWer on page 434 (Channel Power measurements)
FETCh:CTXSpurious on page 436 (TX Spurious Emissions measurements)
FETCh:DAPower on page 439 (Digital Average Power measurements)
FETCh:FM on page 441 (Frequency Modulation measurements)
FETCh:FSTability on page 447 (Frequency Stability measurements)
FETCh:GPOWer on page 453 (Gated Power measurements)
FETCh:HWQuality on page 456 (Handoff Waveform Quality measurements)
FETCh:SAUDio on page 458 (Swept Audio measurements)
FETCh:TFERror on page 464 (TDSO Frame Error Rate measurements)
FETCh:WQUality on page 466 (Waveform Quality/Code Domain power measurements)

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals