98
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_gen_bse_ocns.fm
Configuring the F-OCNS (Forward Orthogonal Channel Noise Simulator)
Configuring the F-OCNS (Forward Orthogonal Channel Noise
Simulator)
The F-OCNS channel parameters are configured using the following call processing subsystem functions and
commands.
Settable F-OCNS Parameters
F-OCNS Walsh code.
Fixed F-OCNS Parameters
F-OCNS level (calculated value based on summation of all other code channels)
OCNS_CCI_SUPPORTED is fixed at 0.
NUM_OCNS is fixed at 1.
F-OCNS Walsh code is fixed at code channel number 80.
GPIB Examples:
Setting F-OCNS Walsh code
OUTPUT 714;"CALL:OCNSOURCE:WALSH CODE5"
would set the F-OCNS Walsh code to 10. See CALL:OCNSource:WALSh[:SELected] .
Querying F-OCNS level
OUTPUT 714;"CALL:OCNSOURCE:LEVEL?"
would query the last F-OCNS calculation. See CALL:OCNSource:LEVel[:SELected]? .
Operating Considerations
F-OCNS is a read-only value.
The F-OCNS value is calculated to provide a summing value that results in a total code channel value
equalling 100% of cell power. If the current desired code channel levels do not allow for a valid OCNS
calculation an error message will be displayed.
Related Topics
CALL:OCNSource
How Do I Change Code Channel Levels?

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals