193
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_gen_inst_states.fm
Preset Descriptions
Status Preset
The STATUS:PRESET command sets the status system as defined in SCPI 1995 Volume 2: Command
Reference section 20.7. All of the enable registers are set to 0, all PTR registers are set to 1, and all NTR
registers are set to 0.
Example
OUTPUT 714;STATUS:PRESET !Presets the STATus subsystem.
Related Topics
SYSTem:PRESet
Table 11. Full Preset Behavior
Function Full Preset Result
Trigger Arm Continuous (manual
operation full preset or
SYSTEM:PRESET2
command)
Trigger Arm Single (remote operation full
preset, *RST command)
Measurement Parameters all set to defaults
Maskable Message Display
State
On (manual operation full
preset)
Maskable Message Display
State
Off (remote operation full
preset)