cdma2000 Measurement Procedures
Measuring Access Probe Power
1. Press the CALL SETUP key.
2. Press the End Call key (
F3).
3. Select the Access Probe Power measurement (see “B. Select measurements.” on page 656).
4. Press the Register Mobile key (
F6). A typical display is shown below.
NOTE The remainder of this procedure lists the steps required to make access probe power
measurements using call limit mode. Call limit mode provides testing of complete access probe
sequences during a call origination.
5. Set up the measurement parameters as needed for your measurement situation (see “B. Set up the
measurement.” on page 658.
6. Press the
More key on the lower left side of the display (Call Control) until 2 of 2 is displayed. Press the
Cell Info key (
F2) key, then the Registration Parameters key.
7. Turn the Timer Based Registration State off.
8. Press the
MEASUREMENT key.
9. Press the
More key on the lower right side of the display (Call Parms) until 2 of 3 is displayed. Press the
Call Limit Mode key (
F9) key and turn call limiting on.
10.Press the Originate Call key (
F3). Wait for the Access Probe call status indication in the Active Cell
field below the measurement window. A typical display is shown below.
11.Turn off call limiting.