EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #320 background imageLoading...
Page #320 background image
320
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_ms_reported_clear.fm
CALL:MS:REPorted:CLEar
CALL:MS:REPorted:CLEar
Diagram Conventions on page 248
CALL:MS:REPorted:CLEar[:ALL]
Function Clears the mobile station (MS) information database.
The following mobile station data is cleared:
Page MS Information,
Status Request Information,
Queriable MS Information (mobile capabilities),
Registration Type,
Mobile Protocol Capability Indication,
MS Called Party Number.
After the MS database is cleared, the test set will re-acquire the mobile station for the Status
Request Information during the next registration.
Programming Example
OUTPUT 714;CALL:MS:REPORTED:CLEAR:ALL
CALL
:CLEar
:MS :REPorted
[:ALL]