Programming a Tx Spurious Emissions Measurement
Programming a Tx Spurious Emissions Measurement
This section provides an example of how to make transmitter spurious emissions measurements via GPIB.
The following program example assumes that the mobile stations antenna output is connected to the RF
IN/OUT connector, and the mobile station has found service.
The basic procedure to measure spurious emissions remotely is:
1. Turn on the mobile station and perform a registration (wait for power-on or send
CALL:REGister[:IMMediate] ).
2. Turn off timer-based registrations (see CALL:REGister:TIMer:STATe ).
3. Set up nominal power, initial power, step size, number of steps, and the maximum number of access probes
for responses (see CALL[:CELL]:APARameter on page 268).
4. Set the trigger arm state to single (see SETup:CTXSpurious:CONTinuous ).
5. Perform a page or mobile station originated call.
6. Initiate the spurious emissions measurement using the INITiate:<measurement mnemonic>[:ON] on
page 478. The <measurement mnemonic> for the cdma2000 transmitter spurious emissions measurement
is CTXS.
7. Use the FETCh:CTXSpurious[:ALL]? command to obtain the measurement results.
Programming Example
10 RE-SAVE "c:\roadhog\programs\tx_spurious.txt"
20 !This program measures conducted spurious emissions from the mobile station
30 !See "Conducted Spurious Emissions" in TIA/EIA-98-D for specifications and standards
40 CLEAR SCREEN
50 A=714
60 !***Specify call parameters***
70 Systype$="DIGITAL2000"
80 Band$="USCELLULAR"
90 Channel=384
100 Sid=1
110 Nid=1
120 Radio_config$="F3R3" !The test standard requires measurements at each radio config
130 Service_opt$="SO2"
140 !
150 !***Specify test parameters***
160 Loss_frequency$="836 MHZ,881 MHZ"
170 Expected_loss$="-2,-2"
180 Nom_pwr=7
190 Init_pwr=15
200 Acc_prb_steps=7
210 Num_step=7
220 Max_rsp_seq=15
230 Acc_prb_size=4
240 Ior=-90
250 Pilot=-7.4

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals