512
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_all_continuous.fm
SETup[:ALL]:CONTinuous
SETup[:ALL]:CONTinuous
Diagram Conventions on page 248
SETup[:ALL]:CONTinuous
Function Sets the trigger arm state for all measurements. Determines whether all measurements will
make one measurement then stop (OFF (single)), or automatically rearm upon completion of one
measurement and repeat the process (ON (continuous)).
This setting overrides previously set trigger arm settings for individual measurements.
Setting Range: ON, OFF
Programming Example
OUTPUT 714;"SETup:ALL:CONTinuous:OFF" !Sets all measurements to single trigger mode.
SETup
[:ALL]
:CONTinuous :ON|:OFF