321
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_ms_reported_cpclass.fm
CALL:MS:REPorted:CPCLass
CALL:MS:REPorted:CPCLass
Diagram Conventions on page 248
CALL:MS:REPorted:CPCLass?
Function Queries the mobile reported power class as part of the station class mark (SCM) for IS-95
mobiles.
Used only in active cell operating mode.
Query Range: 1, 2, 3, or 9.91 E+37
*RST Setting 9.91 E+37
Programming Example
OUTPUT 714;"CALL:MS:REPORTED:PCLASS:SELECTED?" !Returns the MS reported power class.
CALL :MS :REPorted
:CPCLass?
(returns num value)