Configuring the F-SCH (Forward Supplemental Channel)
Configuring the F-SCH (Forward Supplemental Channel)
The F-SCH channel parameters are configured using the following call processing subsystem functions and
commands.
Settable F-SCH Parameters
F-SCH level.
F-SCH state.
F-SCH data rate
F-SCH encoder type (convolutional or turbo)
Fixed F-SCH Parameters
The F-SCH Walsh code is fixed at code channel number 3.
When the F-SCH data rate is below 19.2 kbps, the forward channel convolutional encoder will be selected
regardless of the encoder setting.
GPIB Examples:
Setting the F-SCH level and state (complex command)
OUTPUT 714;"CALL:SCH -15.6 DB"
would set the F-SCH state on and set the level to 15.6 dB. See CALL:SCHannel[:SLEVel][:SELected] .
Setting the F-SCH data rate
OUTPUT 714;"CALL:SCH:DRATE:RCONFIG3 BPS38400"
would set the F-SCH data rate for forward channel radio configuration 3 to 38400. See
CALL:SCHannel:DRATe:RCONfig3 .
Operating Considerations
The F-SCH is available only when the operating mode is set to IS-2000 Test.
The F-SCH channel in forward channel radio configurations 1 and 2 is supported only when operating mode is
set to IS-2000 Test.
Related Topics
CALL:SCHannel
How Do I Change Code Channel Levels?

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals