402
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_sync.fm
CALL:SYNC
CALL:SYNC:LEVel[:SELected]
CALL:SYNC:STATe[:SELected]
Function Sets/queries the forward SYNC (synchronization channel) level.
The optional [:SELected] keyword in this command specifies that the SYNC level being
set/queried applies to the current system type. See CALL[:CELL]:SYSTem[:TYPE] to query or
change the current system type.
Setting Range: 0 to -20 dB
Resolution: 0.01 dB
Query Range: 0 to -20 dB
Resolution: 0.01 dB
*RST Setting -16 dB
Programming Example
OUTPUT 714;"CALL:SYNC:LEVEL -16" !Set the SYNC channel level to -16 dB.
Function Sets/queries the SYNC (synchronization channel) state.
The optional [:SELected] keyword in this command specifies that the SYNC state being
set/queried applies to the current system type. See CALL[:CELL]:SYSTem[:TYPE] to query or
change the current system type.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting ON
Programming Example
OUTPUT 714;"CALL:SYNC:STATE OFF" !Turns off the SYNC channel.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals