594
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_status_operation.fm
STATus:OPERation:
Program Examples - STATus:OPERation:NMRReady:DIGital2000
OUTPUT 714;"STATUS:OPERATION:NMRREADY:DIGITAL2000:EVENT?" !Queries the New Measurement
!Results Ready DIG2000 Event Register.
OUTPUT 714;"STATUS:OPERATION:NMRREADY:DIGITAL2000:CONDITION?" !Queries the New
!Measurement Results Ready DIG2000
!Condition Register.
OUTPUT 714;"STATUS:OPERATION:NMRREADY:DIGITAL2000:ENABLE 8" !Sets New Measurement Results
!Ready DIG2000 Enable Register
!for bit 3.
OUTPUT 714;"STATUS:OPERATION:NMRREADY:DIGITAL2000:NTR 2" !Sets the New Measurement Results
!Ready DIG2000 Negative Transition
!Register for bit 1.
OUTPUT 714;"STATUS:OPERATION:NMRREADY:DIGITAL2000:PTR 4" !Sets the New Measurement
!Results Ready DIG2000 Positive
!Transition Register
!for bit 2
0 1 Extension Bit This bit is a 1 if the measurement has been
completed and has produced new results.
This bit is a zero at power on, after a preset
and while a measurement is in Measuring
States. See Triggering of Measurements on
page 218.
Bit
Number
Binary
Weighting
Condition Description