551
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_gpower.fm
SETup:GPOWer
SETup:GPOWer:CONTinuous
SETup:GPOWer:COUNt[:SNUMber]
SETup:GPOWer:COUNt:NUMBer
Function Sets/queries the trigger arm state for the gated power measurement. The trigger arm state
determines whether the gated power measurement will be made once then stop (0|OFF (single)),
or automatically re-arm upon completion of one measurement and repeat the process (1|ON
(continuous)).
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:GPOWER:CONTINUOUS OFF"
!Sets measurement to single trigger mode.
Function The setting form of this command performs two functions: (1) sets the number of gated power
measurements to be made in a multi-measurement, and (2) sets the multi-measurement STATe
command to 1 (on).
The query form returns the number of gated power multi-measurements to be made when
multi-measurement state is ON.
Setting Range: 1 to 999
Query Range: 1 to 999
*RST Setting Off (10 when STATe = On)
Programming Example
OUTPUT 714;"SETUP:GPOWER:COUNT:SNUMBER 5"
!Turns on multi-measurements and sets the number of measurements to 5.
Function Sets/queries the number of gated power measurements to be made when the multi-measurement
state is ON.
Setting Range: 1 to 999
Query Range: 1 to 999
*RST Setting 10
Programming Example
OUTPUT 714;"SETUP:GPOWER:COUNT:NUMBER 5"
!Sets the number of gated power multi-measurements to 5.