552
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_gpower.fm
SETup:GPOWer
SETup:GPOWer:COUNt:STATe
SETup:GPOWer:TIMeout[:STIMe]
SETup:GPOWer:TIMeout:STATe
Function Sets/queries the gated power multi-measurement state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:GPOWER:COUNT:STATE ON" !Turns on gated power multi-measurements.
Function The setting form of this command performs two functions: (1) sets the gated power
measurements timeout (see Timeouts ) value, and (2) sets the gated power measurements
timeout STATe to 1 (on).
The query form returns the gated power measurements timeout value.
Setting Range: 0.1 to 999.9 seconds
Resolution: 0.1 second
Query Range: 0.1 to 999.9
*RST Setting 10 seconds, on
Programming Example
OUTPUT 714;"SETUP:GPOWER:TIMEOUT:STIME 5 S"
!Sets timeout state to ON and sets the timeout time to 5 seconds.
Function Sets/queries the gated power measurements timeout (see Timeouts ) state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:GPOWER:TIMEOUT:STATE ON"