295
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_call_fchannel.fm
CALL:FCHannel
CALL:FCHannel[:SLEVel][:SELected]
Function The setting form of this command performs two functions: (1) sets the forward fundamental
channel (FCH) level, and (2) turns the FCH state to ON.
The query form of this command returns the FCH level.
The optional [:SELected] keyword in this command specifies that the frequency band being
set/queried applies to the current system type. At this time there is only one system type choice,
DIGital2000. Specifying DIGital2000 will have the same effect as using the optional [:SELected]
keyword.
Setting Range: 0 to -30 dB
Resolution: 0.01 dB
Query Range: 0 to -30 dB
Resolution: 0.01 dB
*RST Setting -15.6 dB
Programming Example
OUTPUT 714;"CALL:FCHANNEL:SLEVEL:DIGITAL2000 -10 dB"
!Sets the forward fundamental channel (FCH) state to ON and sets the level to -10 dB.