CALL:SETup:AVC
CALL:SETup:AVC[:CHANnel][:SELected]
CALL:SETup:AVC[:CHANnel]:CELLular
CALL:SETup:AVC:SATone[:CCODe]
Function Sets/queries the RF channel number of the analog voice channel.
During active cell operating mode, this setting does not take affect until a call is originated on
the AVC or a handoff occurs to a new AVC. This setting takes effect immediately during AVC Test
operating mode.
Setting Range: 1 through 799, 991 through 1023, 9.91 E+37
Query Range: 1 through 799, 991 through 1023, 9.91 E+37
*RST Setting 300
Programming Example
OUTPUT 714;CALL:SETUP:AVC:CHANNEL:SELECTED 799 !Set the AVC to channel 799.
Function Sets/queries the RF channel number of the analog voice channel, specifying the cellular band
(800 MHz).
During active cell operating mode, this setting does not take affect until a call is originated on the
AVC or a handoff occurs to a a new AVC. This setting takes effect immediately during AVC Test
operating mode. PCS band analog operation is currently not supported.
Setting Range: 1 through 799, 991 through 1023, 9.91 E+37
Query Range: 1 through 799, 991 through 1023, 9.91 E+37
*RST Setting 300
Programming Example
OUTPUT 714;CALL:SETUP:AVC:CHANNEL:CELLULAR 799 !Set the AVC to cell band channel 799.
Function Sets/queries the supervisory audio tones color code (frequency).
In active cell mode, this command sets/queries the SAT color code (frequency) to be used on the
next handoff.
In test mode, this command sets/queries the current SAT color code (frequency).
Setting Range: SAT1|SAT2|SAT3
SAT1 = 5970 Hz
SAT2 = 6000 Hz
SAT3 = 6030 Hz
Query Range: SAT1|SAT2|SAT3
*RST Setting SAT2
Programming Example
OUTPUT 714;CALL:SETUP:AVC:SATONE:CCODE SAT1 !Sets the SAT frequency to 5970 Hz.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals