S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_afan.fm
FETCh:AFANalyzer
FETCh:AFANalyzer:SINAD[:AVERage]?
FETCh:AFANalyzer:SINAD:ALL?
FETCh:AFANalyzer:SINAD:MAXimum?
FETCh:AFANalyzer:SINAD:MINimum?
Function Queries the average SINAD measurement.
Query Average SINAD
Range:-99 to 99 dB, 9.91 E+37
Resolution: 0.01 dB
Programming Example
OUTPUT 714;FETCH:AFANALYZER:SINAD:AVERAGE
Function Queries the SINAD measurements.
Measurement results are returned in the following order: minimum SINAD, maximum SINAD,
average SINAD, and standard deviation SINAD.
Query Minimum, maximum, average SINAD:
Range: -99 to 99 dB, 9.91 E+37
Resolution: 0.01 dB
Standard deviation SINAD
Range: -99 to 99 dB, 9.91 E+37
Resolution: 0.001 dB
Programming Example
OUTPUT 714;FETCH:AFANALYZER:SINAD:ALL?
!Returns mininum, maximum, average, and standard deviation SINAD.
Function Queries the maximum SINAD measurements.
Query Range: -99 to 99 dB, 9.91 E+37
Resolution: 0.01 dB
Programming Example
OUTPUT 714;FETCH:AFANALYZER:SINAD:MAXIMUM?
Function Queries the minimum SINAD measurements.
Query Range: -99 to 99 dB, 9.91 E+37
Resolution: 0.01 dB
Programming Example
OUTPUT 714;FETCH:AFANALYZER:SINAD:MINIMUM?