435
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_cpower.fm
FETCh:CPOWer
FETCh:CPOWer:ICOunt?
FETCh:CPOWer:INTegrity?
Function Queries the intermediate count of measurements completed in a multi-measurement.
To set up multi-measurements, see “SETup:CPOWer:COUNt[:SNUMber]” .
Query Range: 0 to 999
Programming Example
OUTPUT 714;"FETCH:CPOWER:ICOUNT?" !Returns the intermediate count of channel power
multi-measurements.
Function Queries the measurement integrity indicator for the last channel power measurement completed.
Zero indicates a normal measurement. See “Integrity Indicator” on page 222 for information
about non-zero integrity results.
Query Range: 0 through 23
Programming Example
OUTPUT 714;"FETCH:CPOWER:INTegrity?"
!Queries the integrity indicator for the channel power results.