S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_fm.fm
FETCh:FM
FETCh:FM:FREQuency:SDEViation?
FETCh:FM:ICOunt?
FETCh:FM:INTegrity?
Function Queries the standard deviation audio frequency.
Query Range: 0 to 99999 Hz, 9.91 E+37
Resolution: 0.01 Hz
Programming Example
OUTPUT 714;FETCH:FM:FREQUENCY:SDEVIATION?
Function Queries the intermediate count of measurements completed in a multi-measurement.
Query Range: 0 to 99
Programming Example
OUTPUT 714;FETCH:FM:ICOUNT?
Function Queries the measurement integrity indicator for the last FM measurement completed.
Zero indicates a normal measurement. See Integrity Indicator on page 222 for information
about non-zero integrity results.
Query Range: 1 through 16
Programming Example
OUTPUT 714;INITiate:FM !Start FM measurement.
OUTPUT 714;FETCH:FM? !Request measurement results.
ENTER 714;Integrity,Fm_dev_dist !Read measurement results.
IF Integrity = 0 THEN !If 0 then measurement was successful.
PRINT FM =;Fm_dev_dist !Measurement results printed if measurement is successful.
ELSE
PRINT Measurement integrity questionable, integrity value = ;Integrity
END IF
END

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals