S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_fetch_gpower.fm
FETCh:GPOWer
FETCh:GPOWer:TRACe?
Function Queries the relative power data points across the display trace. Query returns 5,265
comma-separated values, each corresponding to a point in time across the trace. Data point 222
corresponds with time = 0 on the trace. See How is a Gated Power Measurement Made? for time
values corresponding to all other data points.
Query Range: +10 to -45 dB, 9.91 E+37
Resolution: 0.01
Programming Example
OUTPUT 714;"FETCH:GPOWER:TRACE?"