S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_cfer.fm
SETup:CFERror
SETup:CFERror:CONFidence[:SLEVel]
SETup:CFERror:CONFidence:STATe
SETup:CFERror:CONFidence:LEVel
Function The setting form of this command performs two functions: (1) sets the FER (frame error rate)
measurements confidence level, and (2) sets the FER measurements confidence limit STATe to 1
(on).
The query form of this command returns the FER measurements confidence level setting.
Setting Range: 80 to 99.99 percent
Resolution: 0.01 percent
Query Range: 80 to 99.99
*RST Setting 95 percent, On
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE 95
!sets the FER confidence level to 95 percent
Function Sets/queries the FER (frame error rate) measurement confidence limit testing on/off state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE:STATE ON
Function Sets/queries the FER (frame error rate) measurements confidence limit level.
Setting Range: 80 to 99.99%
Query Range: 80 to 99.99
*RST Setting 95%, on
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE:LEVEL 95
!sets the confidence level to 95%.