EasyManuals Logo

Agilent Technologies 8960 Series Reference Guide

Agilent Technologies 8960 Series
786 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #525 background imageLoading...
Page #525 background image
525
S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_setup_cfer.fm
SETup:CFERror
SETup:CFERror:CONFidence[:SLEVel]
SETup:CFERror:CONFidence:STATe
SETup:CFERror:CONFidence:LEVel
Function The setting form of this command performs two functions: (1) sets the FER (frame error rate)
measurements confidence level, and (2) sets the FER measurements confidence limit STATe to 1
(on).
The query form of this command returns the FER measurements confidence level setting.
Setting Range: 80 to 99.99 percent
Resolution: 0.01 percent
Query Range: 80 to 99.99
*RST Setting 95 percent, On
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE 95
!sets the FER confidence level to 95 percent
Function Sets/queries the FER (frame error rate) measurement confidence limit testing on/off state.
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting 0 (off)
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE:STATE ON
Function Sets/queries the FER (frame error rate) measurements confidence limit level.
Setting Range: 80 to 99.99%
Query Range: 80 to 99.99
*RST Setting 95%, on
Programming Example
OUTPUT 714;"SETUP:CFERROR:CONFIDENCE:LEVEL 95
!sets the confidence level to 95%.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals