S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_status_questionable.fm
STATus:QUEStionable
STATus:QUEStionable:ERRors:TA2000 Register Bit Assignments
The STATus:QUEStionable:ERRors:TA2000 register bits will be used to indicate information about TA2000
related test set device-specific errors (positive error numbers).
Bit
Number
Binary
Weighting
Condition Description
15 32768 Not Used. Defined by SCPI. This bit is always 0.
14 16384 MUI Maskable Message.
13 8192 Reserved for future use. This bit is always 0.
12 4096 Reserved for future use. This bit is always 0.
11 2048 Reserved for future use. This bit is always 0.
10 1024 Reserved for future use. This bit is always 0.
9 512 +900 Errors The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +900 to +999 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
8 256 +800 Errors The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +800 to +899 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
7 128 +700 Errors The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +700 to +799 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
6 64 +600 Errors The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +600 to +699 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.
5 32 +500 Errors The condition bit is pulsed to a 1 and immediately
back to 0 if an error in the +500 to +599 range
occurs. After setting the Positive Transition Filter
and the Negative Transition Filter you can query
the Event Register to find out which of these errors
occurred.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals