S:\Hp8960\E1962B CDMA 2000\Pi Release\Reference Guide\Chapters\cdma2000_hpib_status_std_evnt.fm
Standard Event Status Register
Program Example - Standard Event Status Register
OUTPUT 714;*ESR? !Queries (reads) the Standard Event Status Register.
2 4 Query Error This bit is set to 1 if an error has occurred while
trying to read the test sets output queue. The
following events cause a query error:
An attempt is made to read data from the output
queue when no data is present or is pending.
Data in the output queue has been lost. An
example of this would be an output queue overflow.
1 2 Reserved for future use. This bit will always be 0.
0 1 Operation Complete This bit is set to 1 when the test set has completed all
pending operations and is ready to accept new
commands. This bit is only generated in response to
the *OPC IEEE 488.2 common command.
Bit
Number
Binary
Weighting
Condition Description