EasyManua.ls Logo

Bosch Rexroth ActiveMover - Checking Position of the Calibration Set

Bosch Rexroth ActiveMover
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and repair 167/188
3 842 564 874/2021-04, MIT: ActiveMover, Bosch Rexroth AG
10.5.5 Checking position of the calibration set
NOTICE
An already installed reference WT with correctly adjusted measuring system is
required for the test.
Fig. 125:
B
A
C
E
D
F
1
2
4
3
5
Checking position of the calibration set (1)
Please note:
Perform this check to
ensure that the assembled
calibration set is correctly
positioned.
1. Assemble the reference
system at the rear
and left on the fi xing
element.
2. Fold the knurled screw
upwards.
3. Carefully move the
reference WT under
the reference system.
The WT must not hit the
reference system!
4. Fold the knurled screw
downwards.
5. Slightly press the WT to
the left and tighten the
knurled screw lightly.
Please note:
Depending on the tightening
torque of the knurled screw,
the position may vary by
±5 µm. Always use a uniform
tightening torque.

Table of Contents

Related product manuals