36
Cedex HiRes Analyzer Operator’s Guide
C
C
Step-by-step descriptions for performing measurements via the Measurement dialog box are provided in Section
B, Chapter 1, Setting Up a Single Measurement in the Measurement Dialog Box, in the Cedex HiRes Analyzer
User Training Guide.
For undiluted samples, enter a dilution factor of 1:1. A dilution of 1:2 refers to a mixture of one part cell
suspension and one part dilution medium. A dilution of 1:5 means that one part cell suspension has been
added to four parts dilution medium, etc. You may need to manually dilute samples with very high cell
densities (>1 × 10
7
cells/mL) prior to measuring.
The dilution ratio chosen from the list only relates to the manual dilution of the sample prior to placing the
sample on the system. Dilutions that are an automatic part of the measurement process in the Analyzer are
already taken into account by the Cedex HiRes Software; those calculations are automatically performed and
are not related to entered dilutions.
The Sample Port area indicates which type of sample port should be used for the measurement, based on the
instrument configuration (see the Cedex HiRes Analyzer Operator’s Guide for more information):
c Multi Sampler: use the Multi Sample Tray.
c Single Port: use the single sample port.
c Remote Port: use the Remote Sample Port located at the top of the instrument (see Chapter A, Figure 1).
c Custom Port: Consult with your Cedex administrator for more information about which sample port to
use before running any measurements.
2.3 Interrupting a Measurement
An orange EMERGENCY STOP button is located on the left-hand side of the Measurement dialog box and
can be activated during the measuring procedure to interrupt the analysis process. If the EMERGENCY STOP
button is selected, a message appears asking, Do you want to cancel the process? Select YES to stop and abort
the process immediately. Select NO to put the process on hold.
If YES is selected, the EMERGENCY STOP button changes to read UNLOCK QUEUE. Before more measure-
ments can be run, the system must be unlocked by clicking on this button. If NO is selected, the EMERGENCY
STOP button changes to RESUME. Click on RESUME to resume the measurement process.
Only use the EMERGENCY STOP button in an emergency. Actual measurement data may be lost.