Maintenance and Care
D
D
2 The Multi Sampler DefaultCup
A Cedex Sample Cup is required for performing liquid management routines, such as Prime, LM Shutdown,
calibration, and cleaning routines. The position of the Cedex Sample Cup used for all liquid management
routines is called the DefaultCup position. The standard (default) position for the DefaultCup is position 1 of
the Multi Sample Tray. The DefaultCup position can be changed (see Modifying the Standard Setting for the
Multi Sampler DefaultCup).
The position does not influence measurements; all 20 positions on the Multi Sample Tray can be used for
measurements. However, if a sample is present in a Cedex Sample Cup placed in the DefaultCup position during
a Prime, LM Shutdown, focus adjustment, or any cleaning routines, it can no longer be used for a measurement.
Place an empty Cedex Sample Cup into the position set for the DefaultCup before running LM Shutdown,
focus adjustment, Prime, and all cleaning routines.
The Multi Sampler DefaultCup