EasyManua.ls Logo

Cedex HiRes - Main Application: Counting Cells and Determining Viability; Pre-Measurement Checks and Sample Preparation

Cedex HiRes
218 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
Startup
A
A
2.2 Installing a Reagent Kit
1
Perform an LM Shutdown if the Liquid Management system has been primed
(see Section D, Perform an LM Shutdown).
2
Open the Reagent Kit chamber by pushing the front cover upwards until it clicks into place.
3
Remove the old Reagent Kit with the tray, and dispose of the old reagent bottles.
4
Open the new Cedex HiRes Reagent Kit, and remove the covers from the individual containers.
5
Place the Reagent Kit containers into the appropriate positions in the Reagent Tray:
Detergent Container
Trypan Blue Container
Cleaning Solution Container
Waste Container
Distilled or Deionized Water Container
6
Close the front cover by pulling it all the way down. When the front cover is closed, the capillaries
are automatically correctly positioned in the individual containers.
7
Select the Reagent Kit Status option in the Functions menu of the Cedex Control Center;
the Reagent Kit Status dialog box opens:
8
Click the REPLACE ALL CONTAINERS button in the right-hand corner of the Reagent Kit Status dialog box to
confirm the changing of the Reagent Kit. The Reagent Control display at the bottom right-hand corner of the
Cedex Control Center turns from red to green and indicates the number of runs left.
ccc
Preparing the Cedex HiRes Analyzer for Measurements
Installing a Reagent Kit

Table of Contents