Main Application: Counting Cells and Determining Viability
B
B
5 End of Measurement
A cleaning procedure automatically begins immediately after the scanning ends. The entire Liquid Management
system is rinsed during this procedure (Ultra Fast Clean). Once this procedure is complete, a green bar appears
in the Measurement dialog box, and the message Measurement EndEvent (Measurement) appears in the Event
window on the Cedex Control Center.
To set up a new measurement, click on the NEW MEASUREMENT button. A suggested Reactor ID and Sample
ID will be automatically entered into the Sample ID area by the analysis Software. If a different Reactor ID or
Sample ID is desired, type in the new Reactor ID or Sample ID manually.
When the final measurement in a set of samples is complete, or if there is a longer pause before the next
measurement, perform a Standard Clean. Click on the CLEAN button in the Cedex Control Center, and then
click on Standard Clean. (see Section D, Cleaning Options, in the Cedex HiRes Analyzer Operator’s Guide, for
more information).
Ensure that Cedex Sample Cup from the last measurement remains in the Multi Sample Tray in the DefaultCup
position during the complete cleaning procedure, or attached to the single sample port. Only remove the cup
when a new Cedex Sample Cup is placed in the Multi Sample Tray.
If the system detects impurities in the flow chamber, a message, “Dirt or bubbles detected during Prescan!”
appears, and the Valid field does not have a check in the checkbox. The measurement is completed and results
calculated even if this message appears. If the message appears, check the Prescan image at the end of the
measurement (see Viewing the Prescan Image in the Cedex HiRes Analyzer Operator’s Guide). If the flow
chamber appears to have dirt or impurities, perform a Standard Clean before performing a new measure-
ment (see Carrying out a Cleaning Routine in the Cedex HiRes Analyzer Operator’s Guide).