D Shutting Down the Cedex HiRes Analyzer
Always turn off the Cedex HiRes Analyzer in the following order:
1
Perform an LM Shutdown.
2
Log out of the analysis Software.
3
Shut down the Cedex Control Center.
4
Turn off the Control Unit.
n
1 Perform an LM Shutdown
Perform an LM Shutdown every evening after measurements are finished for the day:
1
Click on the HWM button in the Shortcuts area of the Cedex Control Center to open the Hardware
Management dialog box. Alternatively, select Hardware Management in the Functions menu at the top of the
Cedex Control Center.
2
Click on the LM SHUTDOWN button in the lower-right-hand corner of the dialog box to shut down the
Liquid Management.
The Estimated Total Time area of the Hardware Management dialog box displays the estimated time required
for the LM Shutdown.
3
When everything is shut down, the message “HW started and LM not primed” appears in the Current Status
field.
4
Click on to close the dialog box.
n
Do not shut down the hardware management via the HW SHUTDOWN button, as this may lead to
hardware errors. If the HW SHUTDOWN function is used, the Cedex Software Server and Client must be
restarted again before the hardware management can be restarted (HW STARTUP).
2 Log Out
1
Select Logout under the System menu or click on the LOGOUT button in the Cedex Control Center.
2
If an LM Shutdown has not yet been performed, the message “Do you really want to log out without
performing an LM Shutdown?” appears.
3
Select YES to log out immediately or select NO, perform an LM Shutdown, then log out.
The Software is closed and the Login dialog box reappears.
Logout is possible during a measurement without influencing it. A message indicates that measurements
are in progress.
n
Perform an LM Shutdown