Cedex HiRes Analyzer Operator’s Guide
C
C
Table 4 provides the information required by a person experienced in working with XML files to adapt the XML
example to individual needs.
Customized XML files created for older versions of the Cedex HiRes Software will not work with Software 2.4.
Older files will need to be customized again based on the information provided in Table 4 .
Measurement
Data
XML Tag Syntax Remarks
CupNr CupNumber <CupNumber>
4
</CupNumber>
Workarea TEType Prozesstype,
TEValue
<TEType>
<![CDATA[Prozesstype]]>
</TEType>
<TEValue>
<![CDATA[Standard]]>
</TEValue>
ReactorId RunNumber
<RunNumber>
<![CDATA[Reactor1]]>
</RunNumber>
SampleId Sample PVSNumber <Sample PVSNumber="Sample 1">
CellType TEType Cell Type,
TEValue
<TEType>
<![CDATA[Cell type]]>
</TEType>
<TEValue>
<![CDATA[Std. Size]]>
</TEValue>
CellType and Volume settings
must match.
Precision TEType Precision,
TEValue
<TEType>
<![CDATA[Precision]]>
</TEType>
<TEValue>
<![CDATA[Normal]]>
</TEValue>
Choose the precision settings
(Normal, Minimum, Maximum,
or Superior) as defined in the
Operator’s Manual.
Dilution TEType Dilution,
TEValue
<TEType>
<![CDATA[Dilution]]>
</TEType>
<TEValue>
<![CDATA[1:1]]>
</TEValue>
Dilution= 1 -> Cedex2 Dilution
1:1
Dilution= 2 -> Cedex2 Dilution
1:0.5
Volume Volume
<Volume>300</Volume> CellType and Volume settings
must match. Never leave the
volume empty or set to “0”.
If the volume is empty or set
to “0”, an error will appear.
SampleDrawing CreationDate <CreationDate>
2006-02-10T07:17:11+01:00
</CreationDate>
Comment Remarks
<Remarks>
<![CDATA[Comment1]]>
</Remarks>
Comment field will be combined,
in this example:
Comment1ProductCode:Text1
ProcessCode:Text2
Comment ProductCode
<ProductCode>
<![CDATA[Text1]]>
</ProductCode>
Comment ProcessCode
<ProcessCode>
<![CDATA[Text2]]>
</ProcessCode>
Table 4
The Multi Run Dialog Box
Using the Load Data from XML Option

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cedex HiRes and is the answer not in the manual?

Cedex HiRes Specifications

General IconGeneral
Viability Range0 - 100%
Concentration Range5 x 10^4 to 1 x 10^7 cells/mL
Measurement PrincipleImage analysis
Cell TypesMammalian cells
Measurement ParametersTotal cell count, viable cell count, viability
Measurement TimeTypically < 1 minute
Power Requirements100-240 V, 50/60 Hz
InterfaceUSB and Ethernet
Operating Temperature15-30°C
Operating Humidity20-80% (non-condensing)