Cedex HiRes Analyzer Operator’s Guide
C
C
The table below provides a list of the functions that can be added to user roles, as well the access rights granted
by the selected function.
Function that can be added
to a User Role
Access rights granted by the function
Adjustment Access to the Adjustment and Alignment Verification dialog box;
perform an Automatic Adjustment.
Archiving Access to the Archive and Restore dialog boxes to allow for archiving
measurements and restoring archived measurements.
Audit Trail Access to the Audit Trail dialog box.
Cell Type Access to the Live Operator dialog box and the Cell Type List dialog box;
define, modify, or delete cell types. A user who does not have this function
can still reprocess measurements.
Cleaning Access to the Cleaning dialog box; perform cleaning functions.
Data Exchange Access to the Import to DB and Export to DB dialog boxes to allow for
import or export of complete measurement packages for data exchange
with other instruments.
Database Backup Access to the database backup function.
Database Restore Access to the database restore function.
Diagnosis Access to the Diagnosis dialog box for running various diagnoses of the
system.
Exclude Images from Measurement Allow user to exclude or include individual images in a measurement.
Hardware Management Access and perform functions in the Hardware Management dialog box
hardware management functions: HW Shutdown, Prime, LM Shutdown,
HW Startup.
Instrument Configuration Access and edit information in the Instrument Configuration dialog box.
Liquid Container Configuration Access the Reagent Container Configuration dialog box and set the mini-
mum and maximum liquid levels for reagent and waste containers.
Lock / Unlock Access to the Change System Status dialog box to allow for locking or un-
locking the system (e.g., in case of an SST failure).
Measurement Batch Access the Multi Run dialog box to run batch measurements.
Measurement CTC Access the Cultivation Time Chart dialog box to create and view
Cultivation Time Charts.
Measurement Run Access the Measurement dialog box to run a single measurement.
Measurement View Access the Measurement List dialog box to view a list of measurements in
the database and view results and histograms from past measurements.
Access to the Reprocess function can only be restricted by restricting
access to the Measurement View and Reports functions.
Operational Data Access and reset operational data in the Operational Data dialog box. Print
out information about operational data.
Reagent Kit Status Access and reset the level of reagents in the Reagent Kit Status dialog box.
Report Templates Access the Report Templates dialog box in order to edit report templates.
Reports Access various options for printing reports.
Schedule Access to the Schedule dialog box; define cleaning functions to be
automatically performed at certain intervals.
SST Edit Create and edit SST checkpoints.
SST Run Run SST checkpoints and view SST trending charts.