146
Cedex HiRes Analyzer Operator’s Manual
F
F
Settings in the “Password Settings” tab:
Option Value Description Save last
value
AttemptsToLogin 0 Maximum number of incorrect logins before the user is
locked out.
ExpiresInDays 30 Number of days password is valid [d].
MaxLength 50 Maximum password length.
MinCharacters 1 Minimum number of letters a password must contain.
MinDigits 0 Minimum number of numbers a password must contain.
MinLength 1 Minimum password length.
NeverExpires 1 Indication of whether passwords can become invalid
(see also “ExpiresInDays”). Possible values:
0: passwords become invalid after “ExpiresInDays”
number of days, 1: passwords never become invalid.
OldPassword
ReusedAfter
0 Number of previously used passwords that cannot be
re-used as new passwords.
WarningDays
Before
3 Number of days in advance a user will be warned at
login that a password is about to become invalid.
Settings in the “Reports” tab:
Option Value Description Save last
value
ReportResultPath C:\Cedex2\
reporttmp
Report result .xml path. ×
Report
TemplatePath
C:\Cedex2\
reporttmp
The path to report templates (files extension *.lst).
Settings in the “Focus Finder” tab:
Option Value Description Save last
value
FocusFinder
CoarseStepZ
5 (not adjustable) Coarse focus step for automatic
adjustment.
FocusFinder
FineSamplesZ
20 (not adjustable) Number of samples for fine step
automatic adjustment.
FocusFinder
FineStepZ
5 (not adjustable) Fine focus step for automatic
adjustment.
FocusFinderMaxZ 1250 (not adjustable) Maximum focus position for automatic
adjustment.
FocusFinderMinZ 850 (not adjustable) Minimum focus position for automatic
adjustment.
SaveAutomatic
AlignmentImages
1 Store automatic alignment images in database
(0-False/1-True).
System Options and User Preferences