21
Main Application: Counting Cells and Determining Viability
B
B
The following parameters are listed in both tables:
c
Viable Cell Conc.: Concentration of viable cells [× 10
5
cells/mL].
c
Total Cell Conc.: Total cell concentration [× 10
5
cells/mL].
c
Viability (%): The proportion of viable cells (%) to the total number of cells (viable and dead cells).
c Total Cell Count: The number of all detected objects counted as cells.
c
Avg Compactness: The average variation, in shape, of viable cells from an ideal sphere, with the ideal sphere
shape equaling >1<.
c
Avg Diameter: The average diameter [μm] of viable cells detected.
c Aggregate Rate (%): The percentage of all cells (Total Cell Count) found in aggregates (clusters).
c
Dead Cell Conc.: Dead cell concentration [× 10
5
cells/mL].
c
Viable Cell Count: The number of viable cells counted.
c Dead Cell Count: The number of dead cells counted.
c
Total Object Count: The number of all detected objects counted.
The default concentration value for results is expressed in × 10
5
cells/mL. However, concentrations may
be expressed in × 10
6
cells/mL, depending on the setting in the System Options. Refer to the Cedex HiRes
Analyzer Operator’s Guide for more information.
In the individual result data table, the following three columns precede the above:
c
Image No.: Number of the image.
c
Valid: Indicates whether the Software considered the image quality to be acceptable for analysis.
c Excluded: Checkbox to exclude or include the measurement in the calculation process. This function may
not be available depending on the assigned user rights.
Data Output
Measurement results can be printed out. They can also be exported in various formats for further analysis. For
more information about these functions, see the Cedex HiRes Analyzer Operator’s Guide.
Viewing the Results in the Measurement Dialog Box
Viewing the Results in the Measurement Dialog Box

Table of Contents