AS Autosampler Operator’s Manual
112 Doc. 065051-03 1/08
[388] A NAK command was received from the DSP.
The CPU sent a command to the DSP which the DSP cannot
execute.
Possible Causes: An installed option is not being recognized.
Actions:
Check the INSTALLED OPTIONS screen (see Section C.6.1) to
verify that all optional features are being recognized. A check
mark should appear beside each installed option.
If an installed option is not checked, turn off the AS power and
then turn on the power again and recheck the
INSTALLED
OPTIONS
screen. If the problem persists, contact Dionex.
[389] Timed out waiting for motion complete command from DSP.
The DSP could not complete a motion in the time allowed.
Possible Causes:
Mechanical problem with the autosampler needle arm
A valve is not being actuated
Actions:
Check for an obstruction in the autosampler compartment.
Manually test the autosampler needle arm movement, using the
XYZ TEST screen (see Section C.9.3).
Manually test the sample syringe, the sample prep syringe, and
the injection valve(s), using the
LIQUID CONTROL screen (see
Section C.9.5
).
If the problem persists, contact Dionex.