AS Autosampler Operator’s Manual
120 Doc. 065051-03 1/08
[450] Non-recoverable motion error occurred in Y-axis. Clear any
physical obstructions and then press any key to reset system.
Possible Causes:
Vial obstructing motion
Misaligned sampling needle arm
Actions:
Remove the vial obstruction.
Align the sampling needle arm in the inject port (see Section 5.8).
Turn off the power and check for freedom of movement of all
three axes of the sampling arm.
[451] Recoverable motion error occurred in Z-axis.
Possible Causes: An error occurred during Z-axis movement of
the sampling arm. The DSP was able to recover from the error
and continue the motion.
Actions: Occasional occurrences of this error can be ignored. If
the problem persists, contact Dionex.
[452] Non-recoverable motion error occurred in Z-axis. Clear any
physical obstructions and then press any key to reset system.
Possible Causes:
Vial obstructing motion
Bent sampling needle
Misaligned sampling needle arm
Incorrect X-Y position (not positioned above a vial or well)
Actions:
Remove the vial obstruction.
If the needle is bent, replace the sampling needle assembly (see
Section 5.4
).
Align the sampling needle arm in the inject port (see Section 5.8).