4 Troubleshooting
Doc. 065051-03 1/08 121
Turn off the power and check for freedom of movement of all
three axes of the sampling arm.
Contact Dionex.
[453] Recoverable motion error occurred in inject valve.
Possible Causes:
Valve in wrong position
Actions: Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
[454] Non-recoverable motion error occurred in inject valve. Clear
any physical obstructions and then press any key to reset
system.
Possible Causes: Valve in wrong position
Actions: Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
[455] Recoverable motion error occurred in auxiliary valve.
Possible Causes: Valve in wrong position
Actions: Turn off the AS power and then turn on the power again. If
the message reappears, contact Dionex.
[457]-[474] Syringe reports fatal error:
Syringe reports error:
Possible Causes: Syringe errors can result from various
mechanical problems or from a communication failure between
the syringe and the AS electronics.
Actions:
Check the syringe cable connections.
For fatal errors, press a key to reset the system.
If the error persists, contact Dionex.