D Front Panel Control
Doc. 065051-03 1/08 287
Flushing the Needle
The sample syringe is used to flush the needle with the specified volume
of liquid. If you specify a volume greater than the syringe volume,
multiple flushes are done.
Dispensing from a Reagent Reservoir
DISPENSE is available only if the sample preparation option is installed.
During dispensing, the prep syringe moves a volume of reagent from a
reservoir to a vial.
Volume to Flush
Through Needle
uL
FLUSH
250
Flush Volume
1 to 999 L
μ
1000 uL toRES C #100DISPENSE
Source Reservoir
Destination Vial
Volume to Dispense
Volume
RES A, B,
C, D
1 to 100 (aluminum tray)
1 to 49 (plastic tray)
SMP, S+1, ..., S+9,
WST
1.0 to 99.9 L
100 to 1000 L
μ
μ
Reservoir
Destination Vial

Table of Contents