AS Autosampler Operator’s Manual
288 Doc. 065051-03 1/08
Diluting
DILUTE is available only if the sample preparation option is installed. The
following steps occur during a dilution:
The sample syringe picks up a volume of concentrate from the
concentrate vial and dispenses it to the destination vial.
The prep syringe dispenses a volume of diluent to the destination vial.
The sample syringe mixes the destination vial five times, using 80%
of the volume in the vial.
The sample syringe is flushed with a volume equal to the amount of
concentrate plus one full syringe stroke.
12.3 1234uL
A
uL
+
#S+1 #SMP
DILUTE
Concentrate
Vial
Concentrate Vial
Destination
Vial
Destination Vial
Concentrate
Volume
Concentrate
Volume
Diluent
Volume
Diluent
Volume
Diluent
Reservoir
Diluent
Reservoir
1.0 to 99.9 L
100 to 1000 L
μ
μ
10.0 to
99.9 L
μ
μ
100 to
9999 L
RES A, B
C, D
1 to 100 (aluminum tray)
1 to 49 (plastic tray)
SMP, S+1, ..., S+9
1 to 100
SMP, S+1, ..., S+9
(aluminum tray)
1 to 49 (plastic tray)