AS Autosampler Operator’s Manual
292 Doc. 065051-03 1/08
The table below shows the value ranges for each parameter. See the
following section for details about each parameter.
Controlling the Time Between Injections (Cycle Time)
The CYCLE TIME parameter is an optional feature that can be used to
control the time between injections. See Section 2.7.5
for details.
1. On the
METHOD SETUP screen, position the cursor in the CYCLE TIME
field.
2. If the field displays “--”, press
Select Δ or Select to turn on the
cycle time.
3. Press numeric buttons to enter the desired time between injections.
4. Press
Enter.
Controlling the Needle’s Position in the Vial
The SAMPLE NEEDLE HEIGHT parameter positions the tip of the needle at
the specified distance above the bottom of the vial. This height is always
used for sample injections. It is also the default height for mixing,
pipetting, and dispensing, unless a
NEEDLE operation is specified in the
sample prep steps (see Section D.5.3
).
1. On the
METHOD SETUP screen, position the cursor in the SAMPLE
NEEDLE HEIGHT
field.
2. Press numeric buttons to enter the height setting.
3. Press
Enter.
Parameter Value Range Default
Cycle Time 1 to 240 minutes Off (---)
Sample Needle Height 0 to 46 mm 2 mm*
Tray Temperature
(optional)
4 to 60 °C Off (---)
20 °C (when on)
Wait For Temp Stable Yes/No No
*Sample needle height is measured from the bottom of the vial. Zero mm is closest
to the bottom; 46 mm is closest to the top of the vial.