3 Operation and Maintenance
Doc. 065051-03 1/08 81
Example Simultaneous Injection Program: Two Valves Installed in
an ICS-1000/1500/2000
InjectValve_1.InjectPosition ;Switch the first inject
valve to Inject
InjectValve_2.InjectPosition ;Switch the second
inject valve to Inject
Inject ;Sweep the sample onto
the column
Wait InjectState ;Wait for the injection to
finish
0.000 Pump_InjectValve.LoadPosition ;Switch the first inject
valve to Load
Pump_InjectValve_2.LoadPosition ;Switch the second
inject valve to Load
Load ;Load sample onto the
loop
Wait CycleTimeState ;Wait for the end of the
specified Cycle Time
Pump_InjectValve.InjectPosition ;Switch the first inject
valve to Inject
Pump_InjectValve_2.InjectPosition ;Switch the second
inject valve to Inject
Inject ;Sweep the sample onto
the column
Wait InjectState ;Wait for the injection to
finish