AS Autosampler Operator’s Manual
80 Doc. 065051-03 1/08
Example Program for Simultaneous Mode
The following example Chromeleon or Chromeleon Xpress programs
show the AS commands for sample loading and injecting when the
autosampler is in simultaneous mode. Commands differ slightly,
depending on where the valves are installed.
Example Simultaneous Injection Program: Two Valves Installed in
the AS
Example Simultaneous Injection Program: Two Valves Installed in
an ICS-3000 Detector/Chromatography Module
0.000 Sampler_InjectValve.LoadPosition ;Switch the first inject
valve to Load
Sampler_InjectValve_2.LoadPosition ;Switch the second
inject valve to Load
Load ;Load sample onto the
loop
Wait CycleTimeState ;Wait for the end of the
specified Cycle Time
Sampler_InjectValve.InjectPosition ;Switch the first inject
valve to Inject
Sampler_InjectValve_2.InjectPosition ;Switch the second
inject valve to Inject
Inject ;Sweep the sample onto
the column
Wait InjectState ;Wait for the injection to
finish
0.000 InjectValve_1.LoadPosition ;Switch the first inject
valve to Load
InjectValve_2.LoadPosition ;Switch the second
inject valve to Load
Load ;Load sample onto the
loop
Wait CycleTimeState ;Wait for the end of the
specified Cycle Time