IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-81
PRELIMINARY
8-4-3 Flex HDR (CFE to BEP) Cable Replacement Procedure
8-4-3-1 Tools
Phillips screwdriver.
8-4-3-2 Time Required
10 min
8-4-3-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-4-3-4 Flex HDR (CFE to BEP) Cable Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
2) Open the Front End door assembly and secure the cage door using the support arm as described
in the Front End Door Assembly Opening and Removal Procedure on page 8 - 83.
3.) Unscrew the two securing screws securing the HDR flex cable bracket located on the BEP as
shown in Figure 8-102 below.
4.) Disconnect the HDR flex cable from the Cage as described in the Front End Door Assembly
Opening and Removal Procedure on page 8 - 83.
FRU Part #
Refer to Table 9-20 on page 9-19.
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
Figure 8-102 HDR Flex Cable Bracket Located on BEP