D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-84 Section 8-5 - Electronic Cage Components - Replacement Procedures
PRELIMINARY
3.) Secure the cage door by clipping the cage door securing arm to the device chassis - Figure 8-105.
4.) Disconnect the BE-to-FEPS cable from the Back End, as shown in Figure 8-106.
Figure 8-105 Securing the Cage Door with the Securing Arm
Figure 8-106 Disconnecting BE-to-FEPS Cable from Back End